Data Act for connected medical devices

How manufacturers can design the legally required access to device data so that patient safety, health data and trade secrets remain protected – with a threat model, penetration testing and an integrated management system.

Last updated: October 2026 · Valeri Milke, ISO 27001 & ISO 42001 Lead Auditor

12 Sep 2026Devices placed on the market for the first time after this date need access by design
up to €500,000Maximum fine in Germany for breaching the access-by-design or data provision obligation
> 150Vulnerabilities found by the BSI ManiMed study in ten devices – often in supporting infrastructure
24 h · 72 h · 1 monthReporting stages under the BSI Act for manufacturers classed as important entities

The Data Act for connected medical devices means that providers of connected devices and related services must give the hospital, as the user, access to the data their use generates – on request since 12 September 2025 and, for every device first placed on the market after 12 September 2026, by design. The data must be available by default, securely, free of charge and machine-readable, also to third parties the hospital names. For manufacturers, this creates an additional interface – a “data door” – touching health data, know-how and possibly the integrity of therapy data. This page explains roles, data categories and safeguards, distinguishes the Data Act from the General Data Protection Regulation (GDPR), Medical Device Regulation (MDR), BSI Act and Cyber Resilience Act (CRA), and shows how threat modelling, penetration testing and an integrated management system secure this door.

From entry into force to the design obligation

Verified key dates for data access, supervision and related regulations – tap an entry for details.

Eight key takeaways on the Data Act for connected medical devices

Summary for each key takeaway — expand for the full text.

Which data does the data door open up?

Five data categories of a connected diagnostic and therapy system – select a category. The classification is illustrative and does not replace a case-by-case assessment.

Make available
  • Raw and pre-processed data that are readily available during use, including the metadata.
  • This also includes status, event and error messages as well as usage counters.
  • Provided to the hospital free of charge, in a commonly used and machine-readable format.
Raw dataMetadataError logs
In depth

Building a secure data door: nine chapters from the deadline to the evidence

The chapters lead from the obligations of the Data Act through roles, data categories and safeguards to the threat model, penetration testing and the management system. They are based on a generic reference model that can be applied to many connected diagnostic and therapy systems.

01Starting point

Data access becomes a design obligation

The Data Act moves the question of who is entitled to device data from contract negotiations into product development. For medical devices, this is new – and relevant to security.

The EU Data Act has been in force since 11 January 2024 and has applied since 12 September 2025. Since then, users of connected products have been able to require the data holder to make the data generated by their use available to them or to pass it on to a third party they name – including for devices that were already in the field. In addition, it must be clear before a contract is concluded which data a product generates and how they can be accessed.

The second stage follows on 12 September 2026: connected products and related services placed on the market for the first time after that date must enable access through their design (access by design). Data and metadata are to be available by default, easily, securely, free of charge and in a structured, commonly used, machine-readable format – and, where relevant and technically feasible, directly to the user. In the Commission’s reading, what counts is each individual device, not the device type: even a model that has carried the CE marking for years is affected as soon as a unit is placed on the market for the first time after the cut-off date.

Medical devices are explicitly covered

The regulation’s preamble explicitly mentions medical devices, and according to the Commission’s FAQ the Data Act also applies to products that undergo their own conformity assessment. Before it is placed on the market, a medical device must therefore comply with the Data Act as well as the MDR.

Who is exempt

Micro and small enterprises are exempt from the access obligations, provided they are not linked to a larger company or involved as its partner and do not manufacture or develop on behalf of another company. Medium-sized enterprises only get a one-year grace period: in the first year after reaching that size, and for each product in the first year after it is placed on the market. Irrespective of this, the hospital’s access rights cannot be waived by contract to its detriment; such clauses are not binding on it.

Why this is a security issue

Every additional interface enlarges the attack surface. In the ManiMed study published in 2020 by Germany’s Federal Office for Information Security (BSI), more than 150 vulnerabilities were reported for ten connected medical devices from five categories – often in the supporting infrastructure, rarely in the medical device itself. That is precisely where portals, application programming interfaces and cloud storage for data access are now being built. In Germany, breaches of the design or provision obligation can result in fines of up to €500,000; since 30 May 2026, the competent authority has been the Federal Network Agency.

02Key concepts

Who can demand what: roles and data categories

Who counts as user, data holder or data recipient determines which rights and obligations apply at the data door. In clinical settings, the allocation of roles differs from what intuition suggests.

RoleTypical in clinical settingsRightsLimits and obligations
UserHospital or practice as owner, renter or lessee of the device; recipient of the related serviceFree access to product and service data; sharing with a third party of its own choosingNo use of the data for a competing product; no circumvention of safeguards; GDPR and confidentiality for patient data
Data holderUsually the manufacturer or service provider that technically controls data accessLabelling of trade secrets; justified restriction under the security handbrakeProvision without undue delay; own use of non-personal data only on a contractual basis
Data recipientIndependent service partner, analytics or AI service named by the hospitalData on fair, reasonable and non-discriminatory terms; may owe reasonable compensation – for SMEs, at most the costsPurpose limitation; no development of competing products; no impairment of product safety; no sharing with gatekeepers
Data subjectPatientRights under the GDPR, such as access and data portabilityFor hospital devices, not a user within the meaning of the Data Act

Simplified overview; the actual allocation of roles depends on the contractual and operating model.

Two subtleties matter. Processors within the meaning of the GDPR are not data holders – this applies to cloud providers as well as to the manufacturer, insofar as it processes patient data solely on the hospital’s instructions. And the manufacturer may use non-personal device data for its own purposes, such as product improvement, only on the basis of a contract with the hospital; how far this affects the post-market surveillance analyses required by law needs to be clarified case by case.

Which data are covered

Covered are raw and pre-processed data that are available without disproportionate effort, together with metadata as well as status and error messages. Anything that only arises through additional investment, such as complex proprietary algorithms, is generally excluded unless otherwise agreed – an exemption that follows only from the regulation’s preamble and leaves grey areas. Pseudonymisation or encryption does not turn data into derived data. Exploiting gaps in the device’s protection is not a permissible way to obtain the data.

03Example

The new data door: reference model of a connected diagnostic and therapy system

A deliberately generic example shows where data access arises technically. It can be applied to imaging, infusion technology or ventilation and does not describe any real product.

The reference model consists of four building blocks. An imaging diagnostic device supplies images and measurements. Planning software derives therapy parameters from them, which medical staff review and approve. A therapy device executes the approved plan and records the procedure. Added to this is a manufacturer cloud with separate tenants for each hospital, user management, update delivery and a web portal with an application programming interface (API).

Two design principles are common in such systems: the connection to the cloud is established from inside the hospital, so no inbound access from the internet is needed, and the cloud is decoupled so that it cannot interfere with a therapy in progress. In this model, the data door within the meaning of the Data Act comprises:

  • the web portal for viewing and downloading
  • the API for automated retrieval by the hospital’s systems
  • export paths for large data packages
  • a delegation function through which the hospital names third parties as recipients
  • where applicable, local access on the device or on a gateway

Direct access, where the hospital retrieves data itself, or indirect access, where the manufacturer provides data on request – this design decision has security consequences. Direct access requires a permanently reachable, robustly secured interface, while indirect access shifts risks into manual processes.

Is the cloud service itself a medical device?

That depends on the intended purpose. A service that only stores and forwards data is usually not medical device software; one that performs medical analysis or returns results to the planning stage often is. Companion software without medical device status, such as an installable service or export application, may in individual cases fall under the Cyber Resilience Act; purely web-based services without an associated product, by contrast, are not covered by it.

Data categoryExamples in the reference modelClassification under the Data ActPersonal dataDesigning the provision
Raw measurement and image dataImage series and measurement curves from the diagnostic deviceRaw data, to be made availableHealth dataCommon image or measurement format; prove authenticity with a signature from the capturing device
Planning inputsParameters chosen by physicians, the hospital’s in-house default settingsDeliberately recorded usage data, likely coveredHealth dataOutput read-only; record every medical approval with a timestamp
Calculated therapy parametersOutput of the planning algorithmsLikely derived, grey areaHealth dataStrictest integrity protection; external systems receive no write access whatsoever
Treatment recordSequence, duration, deviations, alarmsProduct data, core scopeHealth dataStore in an audit-proof way; recipients outside the hospital receive pseudonymised extracts
Device statusEvent and error logs, operating hours, software versionsProduct data, to be made availableUsually no personal dataCheck for hidden personal data before provision; label confidential calibration values as trade secrets

Illustrative classification of a generic system; not legal advice – the individual case is decisive.

04Safeguards

The safety and security handbrake and trade secrets

The Data Act provides two mechanisms that allow a manufacturer to limit data access. Both are tied to conditions and cannot serve as a blanket excuse.

The safety and security handbrake

The hospital and the manufacturer can restrict or exclude access to, use or further sharing of data by contract if processing them could undermine security requirements for the connected product laid down by law and thereby have serious adverse effects on the health, safety or security of people. For medical devices, the MDR’s cybersecurity requirements are the obvious candidates – a plausible reading, but not one that is explicitly regulated. The Commission’s draft of non-binding model contractual terms, so far published only in English, includes an option to name the specific security requirement in the contract.

If the manufacturer refuses to share data, it must inform the competent authority – in Germany, the Federal Network Agency, using its forms. The hospital can lodge a complaint there or turn to a dispute settlement body. In the reference model, there would be good grounds for the therapy device not to answer live queries during an ongoing treatment and for safety-critical configurations not to leave the system. By contrast, it would be hard to justify withholding read access to error and event logs from the hospital. In any case, the Data Act does not require write access to planning or control.

Trade secrets in three stages

  • Label and agree: the manufacturer marks trade secrets, including in the metadata, and agrees safeguards with the recipient – such as confidentiality agreements, strict access protocols, standards or codes of conduct.
  • Withhold or suspend: if no agreement is reached, agreed measures are not implemented or confidentiality is breached, the manufacturer may withhold or suspend the sharing of data.
  • Refuse: only in exceptional circumstances and on a case-by-case basis, if the manufacturer demonstrates on the basis of objective criteria that, despite safeguards, it is highly likely to suffer serious economic damage.

In the second and third stages, the decision must be justified in writing and notified to the authority. If secrets are disclosed without protection, their status under the German Trade Secrets Act may also suffer. The threat model and the risk management file provide robust justifications for both handbrakes.

What the Digital Omnibus would change

The Commission’s “Digital Omnibus” amending proposal of November 2025 would add a further ground for refusal under trade secret protection where there is a high risk of secrets leaking to third countries with weaker protection; it leaves the safety and security handbrake unchanged. It has not been adopted: Parliament has not voted, and by mid-October 2026 the Council had not agreed a negotiating mandate – trade secret protection is one of the contested points.

05Boundaries

How data protection, confidentiality and the MDR interact

The Data Act governs who gets access to data. Whether and how those data may be processed, and how the device must be built, are governed by other laws – and they continue to apply.

GDPR and medical confidentiality

Where personal data are involved, data protection law takes precedence; the Data Act does not create a legal basis of its own. If the hospital is the user and the patient the data subject, making data available requires a legal basis under the GDPR and, for health data, one of the exceptions for special categories as well. The hospital usually remains the controller, is responsible for data protection by design and will generally need a data protection impact assessment for connected therapy systems.

On top of this comes the medical duty of confidentiality, backed by criminal law, which the Data Act does not lift. Unless the patient has released the duty or a statutory permission applies, patient data may only flow to service providers to the extent necessary, and only to those who work on the hospital’s behalf and have committed to confidentiality. In our assessment, this does not cover analyses for the manufacturer’s own purposes.

MDR

Software according to the state of the art, including information security, minimum IT security requirements and protection against unauthorised access. A new data interface is a design change that must be assessed and, depending on the class and procedure, agreed in advance with the notified body.

Legacy devices

For legacy devices under the former directives, MDCG 2020-3 treats an additional interoperability channel as a significant change, but not adjustments made to comply with other Union law, provided the benefit-risk ratio is maintained. How this plays out for Data Act interfaces is still open.

BSI Act

Manufacturers of medical devices with at least 50 employees, or with annual turnover and a balance sheet total each exceeding €10 million, generally count as important entities. The act targets corporate IT – from the development environment and signing keys to cloud operations – not product conformity.

Cyber Resilience Act

Medical devices are excluded. Companion software without medical device status may be covered: reporting obligations for actively exploited vulnerabilities and severe security incidents have applied since 11 September 2026, and the remaining obligations apply from 11 December 2027.

European Health Data Space (EHDS)

From 26 March 2029, an additional, authority-mediated route for the secondary use of device data is added. The Data Act explicitly remains unaffected, as does the MDR with its product safety requirements.

06Analysis

Threat model of the data door

A threat model reveals what endangers the new interface – and provides the arguments for justifying safeguards and restrictions.

IEC 81001-5-1 requires a product-specific threat model for health software but does not prescribe a method. The starting point is a data flow diagram of the overall system with its trust boundaries. The Data Act adds a boundary to external recipients selected by the hospital. The boundary around the therapy device remains the one most in need of protection. The MDCG 2019-16 guidance already expects a complete overview of network data flows for connected devices.

The STRIDE method examines every data flow for six threat categories: spoofing, tampering, repudiation, information disclosure, denial of service and elevation of privilege. An attack tree additionally breaks an attacker’s goal – such as “manipulated planning data reach the therapy device” – down into individual steps.

STRIDE categoryPoint of attack at the data doorScenario and consequenceCountermeasureEvidence
SpoofingLogin to portal and APIAttacker uses stolen credentials and acts on behalf of a hospitalMulti-factor authentication; short-lived, bound access tokensTest of login and session management
TamperingHandover of the plan to the therapy devicePlan is altered or assigned to the wrong treatment case – patient harm, possibly permanentSignature and binding to patient, treatment and version; verification on the device before executionIntegrity test; entry in the risk management file under ISO 14971
TamperingUpdate delivery from the cloudA compromised package reaches many hospitals at onceCode signing; rollout in wavesReview of the build and supply chain
RepudiationApprovals and retrievalsA data transfer is later disputed; there is no evidence for the hospital or the supervisory authorityTamper-proof logging of every approval and every retrievalReview of the log chain
Information disclosureObject identifiers in the APIAltered identifiers expose other tenants’ recordsServer-side authorisation check for every objectPenetration test of tenant isolation
Information disclosureNomination of third parties by the hospitalForged recipient nomination or excessive data scope; leakage of health data and trade secretsConfirmation via a second channel; data minimisation; pseudonymisationTest of the approval workflow
Denial of serviceBulk exportWaves of retrievals overload the portal or planningRate limiting; export service decoupled from device operationLoad test
Elevation of privilegeRead accessRead access turns into write or control access; medical approval is bypassedStrict separation of data and control planesTest of function permissions; design rationale

Typical scenarios for a generic diagnostic and therapy system, not findings from a real product.

Assess and transfer to the risk management file

For systems whose cloud connects many hospitals, the cybersecurity guidance issued by the US Food and Drug Administration (FDA) in February 2026 recommends a dedicated architecture view of harm to many patients at once; it rates security risks by exploitability rather than by a probability that can hardly be estimated. Threats with possible hazardous consequences are recorded as causes in the risk management file under ISO 14971. Whenever interfaces or new types of recipients are added, the model must be updated.

07Verification

Penetration tests for data access interfaces

The threat model tells you what can go wrong. The penetration test shows whether the countermeasures actually hold under realistic conditions.

The MDR does not use the term penetration test. The MDCG 2019-16 guidance, however, describes testing as the most important means of security verification and explicitly mentions testing of security functions, fuzzing with malformed inputs, vulnerability scans and penetration tests. IEC 81001-5-1 requires testers to be sufficiently independent. The test scope is derived from attack surfaces, system boundaries and the scenarios of the threat model.

Test priorities at the data door

  • Tenant isolation: using test accounts for two hospitals, the team checks whether other tenants’ records can be retrieved by altering identifiers – the 2023 OWASP API Security Top 10 ranks missing object-level authorisation first.
  • Login and sessions in the portal and API, including the lifetime, binding and reusability of access tokens.
  • Function permissions: can read access be escalated to write or control functions?
  • Delegation to third parties: can an attacker register as a recipient or enlarge the approved data scope?
  • Export paths: are data packages complete and correctly pseudonymised, and are trade secrets labelled?
  • Input handling: how do the API and import functions respond to malformed or maliciously crafted formats (fuzzing)?
  • Logging: are approvals and retrievals recorded completely and in a tamper-proof way?
  • Load limits and decoupling: do planning and device operation remain available during mass retrievals?

Test conditions and report

Testing is carried out as a grey-box test with partial knowledge of the system, using synthetic data sets and in a production-equivalent environment – never on a device in patient use. As the FDA expects, the report states independence, scope, period, methodology and results and justifies any vulnerabilities left open. Findings are rated by exploitability and possible impact on patients; remediation runs as a corrective and preventive action (CAPA) with a retest and is filed in the technical documentation and post-market surveillance. Repeated regularly, the test also demonstrates the effectiveness of the safeguards required by the GDPR and the BSI Act.

08Evidence

Evidence in an integrated management system

The threat model and penetration test only help if their results stay current with every change. This requires a management system that brings all relevant regulations together.

At the data door, the Data Act, the MDR, the GDPR, the BSI Act and the German Trade Secrets Act all meet. If regulatory affairs, information security, data protection and legal maintain separate catalogues and risk scales, contradictions surface in the audit or during an incident at the latest. A natural anchor is the quality management system under ISO 13485, which already governs suppliers, risks, vigilance and changes. Because ISO 13485 does not follow the common harmonised structure of the other ISO management system standards, ISO/IEC 27001, ISO/IEC 27701 and, where applicable, ISO/IEC 42001 are connected via a mapping table.

Shared register

A single inventory of systems, interfaces and data types is the source for the data catalogue under the Data Act, the record of processing activities, the threat model and the list of trade secrets.

Linked risk assessment

A shared risk identifier links patient safety under ISO 14971, information security, data protection and trade secret protection – while each is still assessed against its own appropriate criteria.

Change control board

Every change is checked for MDR relevance, for its impact on the threat model and test scope, and for data protection and Data Act implications.

Suppliers and findings

Service providers for development and cloud operations are qualified against the requirements of all regulations in a single pass; findings from tests and audits run through a shared CAPA process.

Coordinating parallel reporting channels

An incident at the data door can trigger several reporting obligations at once, coordinated by a Product Security Incident Response Team (PSIRT): under the BSI Act, 24 hours, 72 hours and one month to the joint reporting office of the BSI and the Federal Office of Civil Protection and Disaster Assistance (BBK); under the GDPR, 72 hours to the supervisory authority and, as a processor, without undue delay to the hospital; under the MDR, 2, 10 or 15 days if a serious incident has occurred. If the manufacturer then suspends data sharing, this must be notified to the Federal Network Agency; for companion software, reports under the Cyber Resilience Act may be added.

Under the BSI Act, senior management is responsible for implementing and monitoring the security measures. The person responsible for regulatory compliance (PRRC) required by the MDR does not take on the role of the information security officer or the data protection officer; a product security function has proven effective as the link between them. If a hospital names an AI service as a recipient, for example, this makes it possible to clarify in a single pass which data are shared, on what legal and contractual basis, with what residual risk and with what evidence.

09Implementation

Starter plan for manufacturers

Eight steps that manufacturers of connected medical devices should settle before their next shipment – regardless of whether the data door already exists or is still being planned.

  1. Determine applicability

    Record connected products and services, check the SME exemptions and identify the devices placed on the market for the first time after 12 September 2026.

  2. Create a data catalogue

    Record all available product and service data and classify them as data to be made available, derived data, personal data or data requiring confidentiality.

  3. Fulfil information obligations

    Supplement offers, contracts and instructions for use so that the type, scope, format and access route of the data are clear before the contract is concluded.

  4. Prepare contracts

    Draft data-use agreements for the manufacturer’s own analyses as well as clauses on the security handbrake and trade secret protection; the draft model contractual terms serve as guidance.

  5. Design the data door

    Define the access route, separate the data and control planes, and specify tenant isolation, delegation and logging as security requirements.

  6. Update the threat model

    Analyse new trust boundaries and data flows with STRIDE, transfer safety-relevant causes to the risk management file and document the justification for the handbrakes.

  7. Test and demonstrate

    Carry out a penetration test focusing on authorisation, delegation and export paths; close and document findings via CAPA.

  8. Align change and reporting channels

    Discuss the interface with the notified body at an early stage and rehearse the reporting channels to the Federal Network Agency, the BSI, the data protection authority and the vigilance authority in a tabletop exercise.

This page reflects the situation as of October 2026. It does not replace legal advice; questions of interpretation, in particular on where derived data begin and how to handle legacy devices, must be clarified case by case.

FAQ

Frequently asked questions on the Data Act for medical devices

Short answers for manufacturers, regulatory affairs and product security – as of October 2026.

Yes. Connected medical devices are connected products within the meaning of the Data Act, and the Commission’s FAQ confirm that the Data Act also applies to products with their own conformity assessment. A medical device may only be placed on the market if, in addition to the MDR, it meets the applicable Data Act obligations. Micro and small enterprises are exempt, provided they do not belong to a larger group and do not manufacture on behalf of others; medium-sized enterprises only have a one-year grace period.

For every connected device placed on the market for the first time after 12 September 2026. What matters is the individual device, not the device type. Independently of this, hospitals have been able since 12 September 2025 to require the manufacturer to make data available to them or pass it on to third parties – including for devices that were already in the field.

For devices in a hospital or practice, the user is the institution that owns, rents or leases the device or receives the related service. Patients who only encounter such a device during treatment are not users but data subjects under the GDPR; a patient who owns or rents a connected device, for example for home use, can be the user. The data holder is usually the manufacturer, provided it controls data access; insofar as it processes patient data solely on behalf of and on the instructions of the hospital, it acts as a processor and, to that extent, is not a data holder.

Only with a justification and not across the board. The safety and security handbrake allows contractual restrictions if use of the data could undermine statutory security requirements for the device and thereby seriously harm people. If the manufacturer refuses to share data, it must inform the competent authority – in Germany, the Federal Network Agency. A threat model and the risk management file provide the justification.

In principle yes, but only with agreed safeguards. The manufacturer labels its secrets and agrees measures with the recipient, such as confidentiality agreements or access protocols. If no agreement is reached or the measures are breached, it may withhold the data. It may refuse only in exceptional cases where, despite safeguards, it is highly likely to suffer serious economic damage. Both decisions must be justified and notified to the authority.

No. The GDPR takes precedence for personal data, and the Data Act does not create a legal basis of its own. Health data may only be shared if there is a legal basis and an exception for special categories applies. Medical confidentiality continues to apply: unless the patient has released the duty or a statutory permission applies, patient data may only go, to the extent necessary, to service providers who are bound to confidentiality and work on the hospital’s behalf.

A new data interface is a design change that must be assessed and, depending on the class and the conformity assessment procedure, agreed in advance with the notified body. For legacy devices under the former directives, it is still open whether it counts as an adjustment to other Union law or, as a new interoperability channel, ends their transitional status. This should be clarified with the notified body at an early stage.

With a penetration test derived from the threat model. The focus is on tenant isolation and object-level authorisation, plus authentication, function permissions, delegation to third parties, export paths and load limits. Testing is carried out with synthetic data in a production-equivalent environment by suitably independent testers. Findings are remediated through corrective and preventive actions (CAPA), retested and filed in the technical documentation.

Standards & Sources

The content on this page is based on the following publicly available guides and studies.

European Parliament and Council of the EU · 2023

Regulation (EU) 2023/2854 (Data Act)

Art. 2(12)–(14), Arts. 3, 4, 5, 7, 50; Recitals 14, 15, 22: access by design, roles, safety and security handbrake, trade secrets, SME exemptions, dates of application.

European Commission · 2026

Frequently Asked Questions – Data Act, Version 1.4 (22 Jan 2026)

Medical devices in scope, placing on the market per individual product, “safety and security handbrake” and “trade secrets handbrake”, relationship with the GDPR.

European Commission · 2025

Draft Recommendation on non-binding model contractual terms on data access and use, C(2025) 7750

Contract templates between data holder, user and third parties; draft status, so far available in English only.

Federal Republic of Germany · 2026

Data Act Application and Enforcement Act (Datenverordnung-Anwendungs-und-Durchsetzungs-Gesetz, DADG), Federal Law Gazette 2026 I No. 157

Sections 2–4, 15: Federal Network Agency as competent authority, BfDI for data protection, fines of up to €500,000; in force since 30 May 2026 (in German).

Federal Network Agency (Bundesnetzagentur) · 2026

Data Act – information, complaints portal and notification forms

Notifications on the safety and security handbrake and the trade secrets handbrake only via the authority’s forms (in German).

European Parliament (Legislative Observatory) · 2026

Procedure 2025/0360(COD) – Digital Omnibus, proposal COM(2025) 837

Planned amendments to the Data Act and the GDPR; status as of October 2026: not adopted.

European Parliament and Council of the EU · 2016

Regulation (EU) 2016/679 (General Data Protection Regulation, GDPR)

Arts. 6 and 9: legal bases and health data as a special category; Art. 33: notification of personal data breaches.

European Parliament and Council of the EU · 2017

Regulation (EU) 2017/745 on medical devices (MDR)

Annex I Sections 17.2, 17.4, 18.8: software and IT security; Annex IX Section 4.10: changes; Art. 87: vigilance deadlines.

Medical Device Coordination Group (MDCG) · 2020

MDCG 2019-16 Rev. 1 – Guidance on Cybersecurity for medical devices

Overview of network data flows, security testing including penetration tests, joint responsibility.

Medical Device Coordination Group (MDCG) · 2023

MDCG 2020-3 Rev. 1 – Guidance on significant changes (Art. 120 MDR)

New interoperability channels versus adjustments to other Union law for legacy devices.

European Parliament and Council of the EU · 2025

Regulation (EU) 2025/327 on the European Health Data Space (EHDS)

Secondary use of device data from 26 Mar 2029; Data Act and MDR remain unaffected.

European Parliament and Council of the EU · 2024

Regulation (EU) 2024/2847 (Cyber Resilience Act)

Art. 2(2), Art. 71: medical devices excluded; reporting obligations from 11 Sep 2026 relevant for companion software without medical device status.

Federal Republic of Germany · 2025

BSI Act (BSIG) of 2 December 2025, Federal Law Gazette 2025 I No. 301

Annex 2, Sections 28, 32: medical device manufacturers as important entities, reporting stages 24 h, 72 h, one month (in German).

Federal Office for Information Security (BSI) · 2020

ManiMed – Manipulation of Medical Devices, final report

Ten devices from five categories, more than 150 reported vulnerabilities, often in the supporting infrastructure (German edition).

U.S. Food and Drug Administration (FDA) · 2026

Cybersecurity in Medical Devices: Quality Management System Considerations and Content of Premarket Submissions (3 Feb 2026)

Multi-patient harm view, assessment by exploitability, requirements for penetration test reports; non-binding.

OWASP Foundation · 2023

OWASP API Security Top 10 – 2023

API1:2023 Broken Object Level Authorization as the most common API risk; test focus on tenant isolation.

Is your data door ready for the Data Act?

In a no-obligation initial consultation, we review the data catalogue, threat model and test scope of your data access interfaces with you – aligned with the Data Act, the MDR, the GDPR and the BSI Act, as well as MDCG 2019-16 and IEC 81001-5-1.