Book an Appointment

Maintaining connected medical devices

How manufacturers and hospitals keep connected medical devices secure in the field: with an inventory and SBOM, clinically grounded triage, patches as controlled changes, coordinated vulnerability disclosure and aligned reporting channels.

Last updated: October 2026 · Valeri Milke, ISO 27001 & ISO 42001 Lead Auditor

Published inmt|medizintechnik 4.2026

2 · 10 · 15 daysMaximum MDR vigilance deadlines for manufacturers, depending on incident severity
24 h · 72 h · 1 monthReporting stages under NIS2 and the BSI Act for hospitals in scope
> 150Vulnerabilities reported by the BSI ManiMed study in ten connected medical devices (2020)
6 Dec 2025Date from which the new BSI Act with its NIS2 obligations applies

Maintaining connected medical devices means keeping a device not just working once it is on the market, but secure, effective and demonstrably compliant under real conditions of use. Beyond conventional servicing, this requires an up-to-date component inventory with a Software Bill of Materials (SBOM), ongoing situational awareness of vulnerabilities, triage that reflects the clinical context, updates handled as controlled changes, a defined channel for vulnerability reports, and monitoring and incident response. Because an exploited vulnerability can affect diagnosis or therapy, IT security is directly part of patient safety. Responsibility is shared: the manufacturer carries the obligations of the Medical Device Regulation (MDR), while the hospital is responsible for operation and maintenance – in Germany under the Medical Devices Operator Ordinance – and is often also subject to the BSI Act. This page shows how both sides can interlock their processes.

From guidance to reporting obligation

Milestones for the security of connected medical devices in the field – tap an entry for details.

Eight key takeaways on maintenance in the field

Summary for each key takeaway — expand for the full text.

Who contributes what to security in the field?

Joint responsibility under MDCG 2019-16 – select a role.

MDR obligations
  • Monitors the device and its third-party components once on the market and feeds security findings into the risk assessment and technical documentation.
  • Provides updates, security information and instructions, and specifies which interventions the operator may carry out on its own.
  • Assesses incidents for reporting obligations and the need for corrective action, and reports serious incidents within the deadlines.
PMSVigilanceSBOMPSIRT
In depth

Security in the field: nine chapters from servicing to reporting

The chapters follow the life of a vulnerability – from inventory through assessment and change to reporting – and consistently distinguish between what the manufacturer and what the hospital is responsible for. They also make clear whether a statement rests on law, guidance or recommended practice.

01Starting point

From servicing to cyber resilience

For a long time, a device counted as well maintained if it had been inspected, calibrated and was available. For connected medical technology, that is only half the story.

Clinical engineering departments have traditionally planned maintenance around inspection deadlines, manufacturer instructions and downtime costs. For connected devices, that is no longer enough: an ultrasound system, a patient monitor or an infusion pump with a network connection contains an operating system, libraries and interfaces whose security posture changes every day – even when nothing happens to the device itself. A device that passes every electrical safety test can therefore still be an open door for attackers.

The BSI’s ManiMed study showed how real the problem is: Germany’s Federal Office for Information Security examined ten network-capable medical devices from five product categories and reported more than 150 vulnerabilities to the manufacturers over the course of the project. The final report was published in December 2020.

What cyber resilience means in operation

Cyber resilience is the ability to keep operating a device safely despite new vulnerabilities and attacks – or to restrict its use on justified grounds. This adds tasks alongside the service schedule:

  • an up-to-date inventory of all devices, software versions and third-party components,
  • ongoing situational awareness of new vulnerabilities and active attacks,
  • an assessment that links technical severity with clinical consequences,
  • updates and compensating controls handled as documented, approved changes,
  • a channel for vulnerability reports from outside the organisation,
  • monitoring, incident response and reporting within the deadlines.

This page summarises the German-language expert article “Instandhaltung vernetzter Medizinprodukte” (maintenance of connected medical devices) from mt|medizintechnik, issue 4.2026, reorganises it and adds developments up to October 2026.

02Core principle

Security is safety: the clinical view of vulnerabilities

With medical devices, a security incident does not stop at data protection. It can affect the treatment itself.

Beyond confidentiality, integrity and availability, medical devices raise a fourth question: could an attack impair the intended performance or harm people? The MDCG 2019-16 guidance stresses that the security risks of medical devices must not be considered in isolation from their safety risks.

How an IT vulnerability becomes a clinical risk

  • Integrity: tampered readings, images or findings lead to wrong diagnoses.
  • Availability: an encrypted or crashed system delays examinations or procedures.
  • Control: anyone able to change therapy or alarm parameters intervenes directly in treatment.
  • Confidentiality: leaked health data affect data protection and the trust of patients.

This leads to a central rule for assessment in the field: technical severity alone does not determine urgency – what matters is what an attacker could achieve on the specific device and in the specific care process.

Two risk files, one result

Manufacturers carry out risk management according to ISO 14971, which is harmonised under the MDR as EN ISO 14971. Security risks are often assessed in a separate analysis. What counts is the link between the two: every security scenario with a possible impact on patients or users must appear in the safety risk management file, and every safety measure that relies on software needs a security review. In the field, this check is repeated with every new vulnerability and every incident – including the question of whether the benefit-risk ratio needs to be reassessed.

03Framework

Regulations and roles: who carries which obligation

Security in the field is shaped by medical device law, operator law, guidance, standards and NIS2 legislation – each addressing different parties.

MDR: the manufacturer’s obligations

The MDR has applied since 26 May 2021. It requires software developed according to the state of the art with risk management that includes information security, protection against unauthorised access, and minimum requirements for IT networks in the instructions for use. Once a device is on the market, the manufacturer operates a post-market surveillance (PMS) system. From class IIa upwards, this includes a periodic safety update report (PSUR) – at least annually for classes IIb and III and at least every two years for class IIa.

MDCG 2019-16: joint responsibility

The MDCG 2019-16 guidance, in its July 2020 revision, carries these requirements across the entire life cycle. It makes clear that cybersecurity is part of the PMS system and also covers third-party components. At the same time, it describes a joint responsibility: legally, the manufacturer is bound; in practice, suppliers, integrators, healthcare providers and authorities share responsibility as well. Changes to the device, including patches and third-party software, are only intended on the basis of explicit, published manufacturer instructions.

Standards as the state of the art

IEC 81001-5-1

Life cycle standard for health software from December 2021, corrected in December 2025. It covers development and maintenance, extends the process structure of IEC 62304 with security activities and is modelled on IEC 62443-4-1. It is not harmonised under the MDR but is regarded as the state of the art.

IEC 80001-1

In its second edition since September 2021. The standard addresses organisations that integrate health IT systems into their infrastructure and requires risk management before, during and after connection – with the key properties of safety, effectiveness and security.

AAMI TIR57 and SW96

US documents on manufacturers’ security risk management; both explicitly link it to risk management according to ISO 14971.

Medical Devices Operator Ordinance: updates are maintenance

For operators in Germany, the Medical Devices Operator Ordinance (MPBetreibV), in force in its revised version since 20 February 2025, specifies what maintenance includes: alongside servicing, inspection and repair, it explicitly covers the installation of security-relevant software updates – in line with the manufacturer’s instructions. If an update significantly changes how a device is operated, staff must be instructed again. For software classified as a class IIb or III medical device, the ordinance has also required healthcare facilities since 1 August 2025 to carry out IT security reviews in line with recognised technical rules, at least every two years and with a written record.

BSI Act: the hospital’s obligations

Since 6 December 2025, the new BSI Act has transposed the NIS2 Directive into German law. Hospitals belong to the health sector. As a rule, entities with 50 or more employees or with more than EUR 10 million in both turnover and balance sheet total are covered. Those with 250 or more employees or more than EUR 50 million in turnover and EUR 43 million in balance sheet total count as particularly important entities, as do hospitals operating critical facilities – so far defined as 30,000 or more inpatient cases per year – regardless of their size. The obligations explicitly include security in the acquisition, development and maintenance of systems, including vulnerability handling and disclosure, as well as supply chain security.

Cyber Resilience Act and FDA in the wider context

The Cyber Resilience Act does not apply to products covered by the MDR or IVDR. It nonetheless reaches into medical technology: suppliers that sell components with digital elements in their own right have been subject to its reporting obligations since 11 September 2026 and will be subject to its main obligations from 11 December 2027. Internationally, the US Food and Drug Administration (FDA) sets benchmarks: since 29 March 2023, it has required premarket submissions for so-called cyber devices to include, among other things, an SBOM, a plan for handling vulnerabilities including coordinated vulnerability disclosure, and updates on a justified schedule.

04Transparency

Transparency: inventory, SBOM and supply chain

Whether a new vulnerability is relevant can only be answered with an up-to-date list of components – at the manufacturer and in the hospital.

On the day Log4Shell was disclosed, 10 December 2021, many organisations faced the same question: which of our products contain the affected library? For medical devices, this uncertainty is particularly delicate, because hospitals are generally not allowed to scan or modify devices themselves unless the manufacturer provides for it.

What an SBOM delivers

The Software Bill of Materials (SBOM) lists the software components of a product in machine-readable form – with name, version, supplier, hash and dependencies. Two formats have become established: CycloneDX, standardised as ECMA-424 since June 2024, and SPDX, whose version 2.2.1 has been published as ISO/IEC 5962:2021. In Germany, the BSI’s Technical Guideline TR-03183-2 describes the minimum content an SBOM should have.

Obligation or expectation?

In the US, the SBOM for cyber devices is part of the premarket submission. European medical device law contains no explicit SBOM obligation; MDCG 2019-16 cites the bill of materials as an example of security information for operators. In the EU, the expectation follows from the state of the art as set out in IEC 81001-5-1. For supplier components that fall under the Cyber Resilience Act, an SBOM becomes mandatory as part of the technical documentation from 11 December 2027.

From list to situational picture

An SBOM only becomes effective when the manufacturer regenerates it with every build and continuously matches it against vulnerability data. Because a vulnerable library does not automatically mean a vulnerable device, the manufacturer adds statements on actual exploitability (Vulnerability Exploitability eXchange, VEX). Such information is exchanged in machine-readable form via the Common Security Advisory Framework (CSAF), an OASIS standard in version 2.0 since November 2022.

What hospitals should specify in procurement

  • delivery of a machine-readable SBOM with the device and an updated SBOM with every software release,
  • security advisories and VEX statements through a fixed channel, ideally in CSAF format,
  • information on the support period for security updates and on how they are provided,
  • the completed MDS2 form and the minimum requirements for the network environment,
  • a named contact for security questions and incidents.

Risks beyond your own code

Supply chain attacks such as the one on SolarWinds Orion at the end of 2020 or the backdoor in XZ Utils in March 2024 show how malicious code can enter through trusted sources. Countermeasures include verifiable builds, signed artefacts, provenance attestations and a deliberately small set of dependencies; Supply-chain Levels for Software Artifacts (SLSA), currently in version 1.2, provides a tiered framework.

05Assessment

Situational awareness and triage with clinical context

Not every vulnerability is equally urgent. The skill lies in combining technical signals with the clinical role of the device.

Sources for situational awareness

  • security advisories from device manufacturers and component suppliers, ideally machine-readable in CSAF format,
  • the warning and information service of CERT-Bund at the BSI,
  • the Known Exploited Vulnerabilities (KEV) catalogue of the US agency CISA,
  • information sharing within Health-ISAC, the international information sharing community of the health sector,
  • in-house findings from network monitoring, logs and incidents.

Three technical signals

The technical classification relies on three signals: CVSS, in version 4.0 since November 2023, for severity; EPSS for a daily updated forecast of exploitation within the next 30 days; and the KEV catalogue, maintained since November 2021, for confirmed attacks.

The clinical questions

  • Reachability: can the device be accessed from the hospital network, through remote maintenance access or even from the internet?
  • Clinical role: does it control or monitor a therapy, does it deliver findings, is it used directly on the patient?
  • Consequences of tampering: what impact would a loss of integrity or availability have on diagnosis and treatment?
  • Protective measures: are segmentation, access restrictions or monitoring already demonstrably effective?
  • Manufacturer statement: does a VEX statement confirm that the device is affected, or is the component not reachable?

Priority classes as a common language

Manufacturers and hospitals benefit from a small number of clearly described priority classes. The following classification is a proposal; each organisation sets specific response times per product.

ClassTypical characteristicsExpected responseEvidence
EmergencyExploitation confirmed or imminent, device influences therapy or diagnosis, no effective protective measureConvene a crisis team, isolate the device or restrict its function, manufacturer workaround or emergency update, decommissioning as a last resortDecision log, safety assessment, review of reporting obligations
UrgentEasily reachable or noticeable impact on safety or performance, protective measures only partly in placeAccelerated testing and prioritised approval, increased monitoring, rollout planned at short noticeTest report, approval, residual risk
PlannableLimited reachability, effective protective measures in placeUpdate through the regular change process, regular review of protective measuresRisk assessment, schedule, effectiveness check
RoutineBarely reachable, no recognisable link to patient safetyInstall with the next regular software releaseEntry in the inventory and change history

Internal response times are not legal deadlines. The MDR vigilance deadlines and the reporting stages of the BSI Act apply independently (see the chapter on reporting channels).

06Change

Patching as a controlled change – including over the network

An update closes a vulnerability and at the same time changes a validated device. That is why it needs a change process that combines speed with evidence.

Two extremes are risky: reflexively installing every available update, or refusing on principle to change anything on medical technology. MDCG 2019-16 expects operators to have patch processes that roll out security-relevant updates promptly, and an operating environment that allows updates without loss of compatibility. The guidance explicitly provides for updates to be made available outside a field safety corrective action as well.

From advisory to documented update

  1. Establish impact

    Match the advisory against the SBOM, device inventory and VEX statements: which devices, software versions and sites are actually affected?

  2. Assess and classify

    Bring together technical signals, clinical role and existing protective measures, set the priority class and document the safety relevance.

  3. Choose the measure

    Update, configuration change or compensating control such as segmentation, hardening, enhanced monitoring or usage restrictions – aligned with the manufacturer’s instructions.

  4. Test and approve

    The manufacturer verifies and validates the update; the hospital tests it in a realistic environment with its own interfaces and approves it against defined criteria.

  5. Roll out in stages

    Pilot devices first, then the fleet, in agreed maintenance windows and with a prepared way back in case of disruption.

  6. Verify and document

    Check the installation status, confirm functionality, record the residual risk and update the inventory, SBOM and risk management file.

Compensating controls are assessed, approved, time-limited and reviewed in the same way as a patch. From the MDR’s perspective, what counts is evidence that a decision was justified, tested and documented together with its residual risk – not the installation alone.

Updates over the network (OTA)

Over-the-air updates bring fixes to many devices faster, but they create a concentration risk: in the worst case, a faulty or manipulated package reaches the entire fleet. The IETF firmware update architecture, described in RFC 9019 of April 2021, sets out the key principles:

  • a signed manifest describing the origin, compatibility and checksums of the package,
  • verification of authenticity and integrity before installation and again at start-up by a bootloader with a root of trust (secure boot),
  • a fallback image in a separate memory area (A/B scheme), so that a failed update does not disable the device,
  • downgrade protection using sequence numbers, so that attackers cannot install an older, vulnerable but validly signed version,
  • robustness against power failure or loss of connection.

Fallback image and downgrade protection solve different problems: one safeguards availability, the other prevents an attack. On top of this come encrypted transmission, staged rollout waves and telemetry on success rates – and in clinical operation, agreed maintenance windows and clear information for users. If an update significantly changes safety or performance characteristics, a new assessment involving the notified body may be required, depending on the classification.

07Exceptional situation

Zero-days, coordinated vulnerability disclosure and PSIRT

When a vulnerability is known but no update exists yet, preparation is decisive. And for reports to arrive at all, there needs to be an orderly point of entry.

Prepared for the zero-day

A zero-day is a vulnerability for which no patch is available at the time it becomes known. For medical devices, this is particularly difficult: the device has to keep supporting patient care, an approved update is missing, and unilateral interventions by the hospital are ruled out. Four precautions have proven effective:

  • a documented escalation path that defines who decides in the first hours – product ownership, security, hospital IT, clinical engineering and regulatory affairs,
  • immediate measures agreed in advance, such as network isolation, disabling services that are not needed and targeted monitoring for signs of attack,
  • prepared and clinically validated fallback configurations for safety-critical device classes,
  • templates for safety communications to users (field safety notices) that are technically correct and clinically understandable.

Under the MDR, a field safety corrective action must in principle be reported to the authority before it is carried out; only in urgent cases may the report follow afterwards. If a vulnerability is already being actively exploited, the situation becomes even more serious – technically as well as from a regulatory perspective.

Coordinated vulnerability disclosure

Many vulnerabilities are discovered by researchers, penetration testers, integrators or users. Without a visible reporting channel, such reports are left unanswered or become public in an uncoordinated way. Coordinated vulnerability disclosure (CVD) brings order to the process: the report is received, verified and fixed, and the vulnerability is then disclosed jointly. The framework is set by ISO/IEC 29147:2018 for receiving reports and publishing advisories and by ISO/IEC 30111:2019 for internal handling.

Point of contact

A security.txt file according to RFC 9116 of April 2022 under /.well-known/ on the website, a policy for reporters and a contact that supports encryption.

Handling

A Product Security Incident Response Team (PSIRT) with defined response times; FIRST’s PSIRT Services Framework offers guidance.

Identification

Unique CVE identifiers, assigned through the manufacturer’s own CVE Numbering Authority (CNA) or in cooperation with an existing CNA.

Publication

Security advisories listing affected versions, measures and VEX status, ideally machine-readable in CSAF format.

What hospitals should check

Without a working CVD process, a manufacturer’s ability to respond is hard to gauge. Telling indicators are a published policy, a reachable PSIRT and a traceable history of security advisories.

08Incident

Monitoring, incident response and parallel reporting channels

Often only hours lie between the first anomaly and a reporting obligation. Clarifying who reports what only once an incident is under way wastes valuable time.

Monitoring beyond availability

Monitoring that only checks whether a device is reachable will not detect attacks. Meaningful indicators include unusual connections, repeated failed logins, unexpected configuration changes and failed integrity checks. Because no additional software may be installed on many medical devices, passive network monitoring in segmented device zones is a suitable approach.

Incident response across organisational boundaries

Anomalies must be turned into an actionable process – with roles, escalation criteria and interfaces to clinical engineering, hospital management, data protection and the manufacturer. MDCG 2019-16 explicitly expects operators to inform the manufacturer without delay of any suspected security incident. At the manufacturer, the PSIRT coordinates analysis, prioritisation and remediation and supports communication with CERTs, information sharing communities and authorities.

Four reporting channels compared

Reporting channelWho reportsTriggerDeadlinesRecipient
MDR vigilanceManufacturerSerious incident, including one with a digital cause, or field safety corrective actionWithout delay; at the latest 2 days for a serious public health threat, 10 days for death or an unanticipated serious deterioration in health, otherwise 15 daysCompetent authority, in Germany usually BfArM
User report (Germany)Operators and users, such as the hospitalSuspected serious incident involving a medical deviceWithout delayCompetent federal authority, usually BfArM
BSI Act (NIS2)Hospital as a particularly important or important entitySignificant security incidentEarly warning within 24 hours, incident notification with an initial assessment within 72 hours, final report no later than one month after the notificationBSI
GDPRController, such as the hospitalPersonal data breach posing a risk to data subjectsWithout undue delay, where feasible within 72 hours; delays must be justifiedData protection supervisory authority

The deadlines run independently of one another, generally from the point of awareness; the final report under the BSI Act is counted from the 72-hour notification. In addition, manufacturers submit trend reports when non-serious incidents show a statistically significant increase.

For MDR vigilance, it makes no difference whether an event stems from an attack, a vulnerability or a software bug. What matters is whether the device’s safety or performance was or could have been impaired. MDCG 2019-16 contains examples that distinguish incidents from serious incidents; for coding, the International Medical Device Regulators Forum (IMDRF) provides problem codes such as “Software Security Vulnerability”. If it is unclear whether a report is required, report within the deadline – an initially incomplete report is permitted.

What the revision proposal would change

On 16 December 2025, the Commission proposed revising the MDR and IVDR. Under the proposal, manufacturers would report actively exploited vulnerabilities and severe security incidents within 30 days via the European database EUDAMED to the competent CSIRTs and to ENISA; in addition, the general vigilance deadline for other serious incidents would be extended from 15 to 30 days. The proposal has not yet been adopted. Until then, the current deadlines apply.

09Implementation

Target state and starter plan for maintenance in the field

Maintenance in the field only becomes effective when security, quality, regulatory affairs and clinical operations share a common governance process.

Post-market surveillance, vulnerability management and incident response depend on one another: new risks only become visible if security intelligence feeds into post-market surveillance; priorities are only right if the safety assessment is part of the process; deadlines are only met if regulatory affairs is involved early on. The target state links these strands into a cycle with clear ownership.

Transparency

A complete inventory of devices, software versions and third-party components, based on up-to-date SBOMs.

Situational awareness

Ongoing analysis of manufacturer advisories, CERT-Bund, KEV and in-house observations.

Decision

Risk-based choice between update, OTA wave, compensating control or restricted use – with a safety assessment.

Report intake

A working CVD channel and a PSIRT that handles reports reliably.

Evidence

PMS results, an updated risk management file, change history and, where applicable, reports in the technical documentation.

Accountability

Defined roles for product, quality, security, service, regulatory affairs and clinical operations, with review cycles and escalation criteria.

Starter plan for the first months

  • Inventory and owners: record all connected medical devices with software version, network connection, manufacturer contact and end of support.
  • Close gaps in manufacturer information: request SBOMs, MDS2 forms, security advisories and patch instructions, and anchor them in contracts.
  • Define priority classes and internal response times that combine technical and clinical criteria.
  • Set up a change process for updates and compensating controls with testing, approval and documentation steps, and align it with the maintenance and inspection duties under the Operator Ordinance.
  • Create and rehearse a reporting matrix for MDR vigilance, user reports, the BSI Act and data protection, with templates, responsibilities and deputies.
  • Introduce KPIs, such as the share of devices with an up-to-date SBOM, lead time per priority class and the number of time-limited compensating controls.

Organisations that connect these building blocks early respond faster, can justify their decisions and strengthen trust in digital medical technology. A securely developed device only stays secure if it is maintained just as carefully in the field.

The information on this page does not replace legal advice. Which obligations apply in an individual case depends on the classification, role and size of the organisation.

Expert article · mt|medizintechnik

Maintenance of connected medical devices

In this German-language journal article, Valeri Milke, CEO of VamiSec GmbH and Lead Auditor for ISO/IEC 27001 and ISO/IEC 42001, explains why vulnerability and patch management for connected medical technology are a matter of patient safety, and how manufacturers and operators can organise maintenance as one continuous process. This page summarises the article, restructures it and adds recent developments such as the new Medical Devices Operator Ordinance, the new BSI Act and the MDR revision proposal.

Issue4.2026
PublisherTÜV Media
FocusMaintenance
Scope5 figures, 1 table
  • A device that works on paper can still pose an unacceptable security risk.
  • Vulnerabilities are only prioritised correctly with clinical context.
  • Patches and OTA updates are controlled changes with testing, approval and evidence.
  • MDR vigilance, NIS2 and data protection call for coordinated reporting processes running in parallel.

The full article was published in the print edition and in the e-paper for subscribers.

FAQ

Frequently asked questions on maintaining connected medical devices

Short answers for manufacturers, clinical engineering and hospital IT – as of October 2026.

Besides servicing and inspections, it involves an up-to-date inventory with software versions and SBOMs, continuous tracking of new vulnerabilities, an assessment that includes the clinical context, updates and compensating controls as controlled changes, a channel for external vulnerability reports, and monitoring, incident response and reporting. At the manufacturer, the results feed into post-market surveillance and risk management; in the hospital, they feed into the operation of medical technology and information security management.

Yes, provided the manufacturer allows it. In Germany, the Medical Devices Operator Ordinance even counts the installation of security-relevant software updates as part of the operator’s maintenance duty, always in line with the manufacturer’s instructions. MDCG 2019-16 only provides for changes to the device, including patches and third-party software, on the basis of published manufacturer instructions. Once the manufacturer releases an update, the hospital installs it promptly through an orderly process with testing, approval and documentation. Without a release, compensating controls such as segmentation or restricted use remain the option.

European medical device law currently contains no explicit SBOM obligation. MDCG 2019-16 cites the bill of materials as an example of security information for operators, and IEC 81001-5-1 shapes the state of the art. In the US, the FDA has required an SBOM in premarket submissions for cyber devices since March 2023. Hospitals should make it a contractual requirement in procurement, and manufacturers should provide it in machine-readable form in CycloneDX or SPDX.

No. The Cyber Resilience Act excludes products covered by the MDR or IVDR, because medical device law sets its own security requirements. It does, however, have an indirect effect: suppliers that sell components with digital elements in their own right have been subject to its reporting obligations since 11 September 2026 and will be subject to its other obligations, for example on SBOMs, from 11 December 2027. The MDR revision proposal also adopts terminology from the CRA.

What must be reported is a serious incident: a malfunction or deterioration in performance – including one caused by an exploited vulnerability – that led or might have led to death, a serious deterioration in health or a serious public health threat. The maximum deadlines are 2, 10 or 15 days. A known vulnerability on its own does not trigger a report, but its potential consequences must be assessed and documented. MDCG 2019-16 contains examples that help draw the line.

If the hospital falls under the BSI Act as a particularly important or important entity, it reports a significant security incident to the BSI with an early warning within 24 hours, a notification with an initial assessment within 72 hours and a final report within one month of that notification. If personal data are affected, the data protection authority must be informed within 72 hours where feasible. A suspected serious incident involving a medical device goes without delay to the competent federal authority, usually BfArM, with the manufacturer involved.

By combining technical signals with clinical context. CVSS describes severity, EPSS the probability of exploitation in the next 30 days, and the KEV catalogue confirms active attacks. Added to this are the device’s reachability, its proximity to the patient, possible consequences for diagnosis or therapy, effective protective measures and the manufacturer’s VEX statements. The result is a small number of priority classes with response times defined per product.

Coordinated vulnerability disclosure is a structured procedure through which a manufacturer receives, verifies and fixes external vulnerability reports and then discloses them in a coordinated way – according to ISO/IEC 29147 and ISO/IEC 30111. For hospitals, it indicates how quickly a manufacturer responds to critical vulnerabilities. Worth checking are a published policy, a security.txt contact, a reachable PSIRT and a traceable history of security advisories.

Standards & Sources

The content on this page is based on the following publicly available guides and studies.

mt|medizintechnik · TÜV Media GmbH · 2026

Valeri Milke: Instandhaltung vernetzter Medizinprodukte (Maintenance of connected medical devices), issue 4.2026

Journal article on which this page is based (issue focus: maintenance; in German).

Official Journal of the EU / EUR-Lex · 2017

Regulation (EU) 2017/745 on medical devices (MDR)

Annex I Sections 17.2, 17.4, 18.8, 23.4; Arts. 83–89: PMS, PSUR, vigilance deadlines, trend reporting, follow-up actions.

MDCG / European Commission · 2020

MDCG 2019-16 Rev. 1 – Guidance on Cybersecurity for medical devices

Joint responsibility, obligations of operators and integrators, PMS and vigilance, IMDRF codes.

European Commission · 2025

COM(2025) 1023 – Proposal amending the MDR and IVDR (2025/0404(COD))

Proposal of 16 Dec 2025: new Art. 87a MDR (30-day reporting to CSIRTs/ENISA), vigilance deadline 15 → 30 days; not yet adopted.

IEC · 2021

IEC 81001-5-1:2021 – Health software and health IT systems safety, effectiveness and security – Part 5-1: Security – Activities in the product life cycle

Security activities in development and maintenance; corrected version 12/2025; not harmonised under the MDR (as of June 2026).

IEC · 2021

IEC 80001-1:2021 – Application of risk management for IT-networks incorporating medical devices – Part 1

Risk management by the healthcare delivery organisation for medical IT networks.

U.S. Food and Drug Administration · 2026

Cybersecurity in Medical Devices: Quality Management System Considerations and Content of Premarket Submissions

Final since 3 Feb 2026; Section 524B FD&C Act: SBOM, postmarket plan with CVD, update cycles.

BSI · 2025

Technical Guideline TR-03183 Part 2: Software Bill of Materials (SBOM), version 2.1.0

Minimum SBOM content; CycloneDX 1.6 or later or SPDX 3.0.1 or later.

OASIS Open · 2022

Common Security Advisory Framework (CSAF) Version 2.0

Machine-readable security advisories including the VEX profile.

FIRST · 2026

Exploit Prediction Scoring System (EPSS)

Daily estimate of the probability of exploitation per CVE over the next 30 days.

CISA · 2026

Known Exploited Vulnerabilities Catalog

Catalogue of vulnerabilities with confirmed exploitation since November 2021; now governed by BOD 26-04.

Federal Republic of Germany · 2021

Medical Devices User Reporting and Information Ordinance (Medizinprodukte-Anwendermelde- und Informationsverordnung, MPAMIV)

Section 3: operators and users report suspected serious incidents without delay to the competent federal authority.

IETF · 2021

RFC 9019 – A Firmware Update Architecture for Internet of Things

Signed manifests, secure boot, downgrade protection, fallback image.

Federal Republic of Germany · 2025

Ordinance on the Operation and Use of Medical Devices (Medical Devices Operator Ordinance, MPBetreibV), 2025 revision

In force since 20 Feb 2025; Section 7: maintenance including security-relevant software updates; Section 4(3): instruction after updates; Section 17: IT security reviews (applicable from 1 Aug 2025).

Federal Republic of Germany · 2025

Act on the Federal Office for Information Security (BSIG) as amended by the NIS2 Implementation Act (NIS2UmsuCG)

Transposes Directive (EU) 2022/2555 (NIS2); in force since 6 Dec 2025; Sections 28, 30, 32 and Annex 1: entity categories, measures, reporting stages 24 h / 72 h / one month.

BSI · 2020

ManiMed – Manipulation of Medical Devices, final report

Ten connected medical devices from five categories, more than 150 reported vulnerabilities.

How robust is your maintenance in the field?

In a no-obligation initial consultation, we review your process – from device inventory through triage and patch approval to the reporting matrix – against the requirements of the MDR, the Medical Devices Operator Ordinance and the BSI Act, as well as IEC 80001-1, IEC 81001-5-1 and the MDCG 2019-16 guidance.